Background

Galacto-oligosaccharides (GOS) are natural functional oligosaccharides, typically built from a galactose or glucose core carrying 1–7 galactosyl units, with degrees of polymerisation ranging from 2 to 8. In the human gut they are used by probiotics such as bifidobacteria and lactobacilli, promoting their growth and proliferation. GOS is abundant in human milk, effectively supports bifidobacteria proliferation in infants, and offers further nutritional benefits including improved mineral absorption, constipation prevention and immune support. As a result, GOS is added to many infant formula products.

Several methods exist for GOS determination: thin-layer chromatography, gas chromatography, and high-performance liquid chromatography (HPLC-ELSD, HPLC-RID and HPLC-MS). Ion chromatography is also used, and high-performance ion chromatography is the main approach for GOS in milk powder. Thin-layer chromatography requires home-made plates together with prepared developing and colouring reagents, which makes it laborious and poorly quantitative. HPLC sample preparation is also complex, usually needs several standards, and the mass spectrometric detectors involved raise the cost of analysis.

This case applies ion chromatography to the determination of GOS in samples, using a Wayeal ion chromatograph with an amperometric detector and sodium hydroxide/water eluent. Sample preparation is simple, recovery is good and sensitivity is high.

Instrument Configuration and Test Conditions

ItemDescription
Ion chromatographIC6300 intelligent ion chromatograph
Analytical columnPA20, 3 × 250 mm
DetectorAmperometric, gold electrode (AAA amino acids, sugars, water-soluble vitamins waveform)
Consumables0.22 µm aqueous syringe filter, 2 mL disposable syringe, 0.1 mg balance
ReagentsGalactose, glucose and lactose standards (1000 mg/L each); KH2PO4/K2HPO4; HPLC-grade acetonitrile; 50% NaOH; anhydrous sodium acetate; galactosidase (~150,000 U/g)
ParameterSetting
Eluent200 mmol/L NaOH + sodium acetate, gradient
Flow rate0.4 mL/min
Run time70 min
Injection volumeFull-loop injection, 20 µL
Column temperature30 °C
Detection cell temperature30 °C

Sample Preparation

An appropriate amount of sample is prepared according to GB 1903.27-2022, the national food safety standard for the food nutritional fortifier galacto-oligosaccharide. The extract is diluted to a suitable factor, filtered through a 0.22 µm aqueous filter and injected. A blank solution is run in parallel to confirm a clean baseline before quantification.

Results

Linearity: five-level calibration series were prepared for galactose, glucose and lactose. All three sugars showed correlation coefficients above 0.999, confirming good linearity over the working range.

SugarLinear equationR
Galactosey = 4107.57500x + 438.010300.99938
Glucosey = 4956.41900x + 762.555200.99903
Lactosey = 2942.84300x − 506.009900.99994

Repeatability: a standard mixture was injected six times. Retention times were highly stable (galactose 7.144 min, glucose 7.994 min, lactose 17.322 min) with retention-time RSD of 0.435%, 0.503% and 0.758%, and peak-area RSD of 0.790%, 0.708% and 0.729% respectively. Calculated concentrations averaged 1.050, 1.001 and 1.082 mg/L against a nominal 1 mg/L, with RSD of 0.861%, 0.801% and 0.606%.

ParameterGalactoseGlucoseLactose
Mean retention time (min)7.1447.99417.322
Retention time RSD (%)0.4350.5030.758
Peak area RSD (%)0.7900.7080.729
Mean concentration (mg/L)1.0501.0011.082

Sample testing: three GOS samples (A0, A1 and A2) were prepared as described and quantified by external standard calibration, with consistent chromatographic profiles across the series.

Conclusion

Using a Wayeal IC6300 intelligent ion chromatograph with amperometric detection, GOS in samples can be determined qualitatively and quantitatively after simple sample preparation, chromatographic separation and external standard calibration. The method is easy to operate, repeatable and sufficiently sensitive and accurate — a practical alternative to thin-layer, gas and HPLC approaches for GOS content control in milk powder and related products.